PROBLEM CAUSE(s)
REMEDY
The wrench does not
click.
The pawl is not in the pawl
slot.
OR
There is no pressure on the
pawl.
Disassemble wrench and
add washers to increase
pressure on pawl.
NOTE
When placing the
ratchet/head drive into the
tube to insert the pivot pin,
the drive pivot hole must be
pushed into the tube to align
the holes.
If there is no pressures
remove parts and add more
washers to increase pressure
on the pawl.
The wrench only clicks
in one direction or can
only be calibrated in one
direction.
The head drive and/or
secondary arm (if used) are not
in centerline.
Replace the head drive
and/or secondary arm (if
used).
The pawl double clicks.
Check
and
replace
as
needed head and/or
secondary arm (if used).
Check and replace the pawl,
as needed.