9 sleep mode configuration – Salter Brecknell 7010SB User Manual
Page 6

SCALE CALIBRATION INSTRUCTIONS
Periodically your scale may require calibration to ensure proper weighing. You
will need a 10-pound weight to calibrate the scale. Scale calibration is
accomplished, when needed, by the following instructions:
Function
Scale Response
1. Remove all weight from the scale
2. Press and hold the “ON/TARE” key and
press the “lb-oz/grams” key once . . . . . . . . . . . . . . . . . “SLP0/SLP1”
3. Press the “ON/TARE” once to accept the
sleep mode setting and enter into calibration mode. . . .”64XX – 10XXX”
4. Press the “ON/TARE” key once
to zero the scale. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .“0”
5. Place 10 lbs on the scale . . . . . . . . . . . . . . . . . . . . . . . ..“8,9XX - 10,000”
6. Press “ON/TARE” key once
To accept calibration weight . . . . . . . . . . . . . . . . . . . . . . .“CAL” then * “Err”
7. Press the “OFF” key to exit calibration mode.
8. Remove the weight from the scale, press the “ON/TARE” key to continue the
normal weighing mode.
* Note: The “Err” displayed after completing calibration does not indicate a
calibration error. Since, the scale only weighs up-to 9 lbs. 15.5 oz., the ten
pounds on the scale causes the displayed message “Err”.