Scale calibration instructions – Salter Brecknell 7010SB User Manual
Page 5

Error Conditions – Shown in the scales display
Below are the error conditions that may appear on the scale display.
“- - - -“
Under Capacity
When the scale weight is below zero the scale will display all dashes (----) and
transmit a status for this condition. This situation will likely occur after the use of
tare when the tare weight is removed. To clear this indication, press the OFF
key, then the ON key and the scale will re-zero.
“Err” Over Capacity -
When the scale is in an overload condition the scale will also display “OvEr” and
transmit a status for this condition. In this case the weight applied to the scale is
above the 10-pound limit; remove the weight from the scale to clear this
indication.
“Lo”
Low
Battery-
When “Lo” is displayed replace the battery.
ErrL, ErrH, 8888:
The scale is damaged. If within warranty, return the scale to the factory.
TROUBLE SHOOTING
Symptom
Recommended
Action
Not weighing correctly OR
Inaccurate weight. . . . . . . . . . . . . . . . . Press OFF key, then the ON key
before placing item on the scale.
OR
The scale may need re-calibration. See
calibration instructions.
Weight fluctuates . . . . . . . . . . . . . . . . . . Press OFF key, then the ON key
before placing item on the scale.
OR
Ensure scale is placed in a stable, level
environment.
PN# 7010SB-01 REV. A Page
5
of
7