Calibration mode – Salter Brecknell 7620U User Manual
Page 17

Model 7620U Bench Scale User’s Manual
17
Calibration Mode
10. Press the F1 key.
Parity selection is stored.
11. Press the ZERO key…
“PROT” is displayed.
12. Press the F1 key…
The current setting is displayed. Use the
ZERO key to toggle between the four
choices: “NCI STD, H-100, 7611, NCI
SMA.”
13. Press the F1 key…
Protocol selection is stored.
14. When finished configuring your scale, press
the ZERO key until the display shows
“DONE,” then press the F1 key.
Or, move Switch 1 to CLOSED position for
normal weighing mode.
The Calibration (CAL) Mode menu lets you
calibrate your scale. The items in the Calibration
menu are as follows:
Capacity (5-30, 6-30, 6-35, 5-50)
Select the capacity of your scale.
Follow these steps to calibrate your scale. Refer
to Figure 2.
1. From normal weighing mode, move
Switch1 to the Menu Mode or OPEN posi-
tion…
“DIAG” is displayed. Press the ZERO
key until “CAL” is displayed.
Calibration (CAL) Mode
Warning! Entering into
this mode can erase
the calibration already
saved. You need
approved calibration
weights to use
calibration mode.
Step-by-Step
Instructions for
CAL Mode