Setup menu mode, Setup menu chart, Continued – Salter Brecknell PC150 User Manual
Page 8
8
Revision 1.2
Setup Menu Mode /
Continued
To place the unit in Setup menu mode:
1. Turn the power off to the unit.
2. Locate the calibration switch on
the scale’s back panel and toggle
to the opposite position.
3. While holding down both the
ZERO
and
keys, turn the
power back on.
4. When the display shows "
F1
",
the unit is in Setup Menu mode,
and you can release the two
keys. Shown at right are the di-
rectional and SET key assign-
ments
SET
SETUP MENU CHART
F1
Display Mode
F6
Zero Range
F9
Calib. Unit.
0
1
F11
Display Check
F12
Calibration
Press ZERO
key to begin
Press ZERO
key to begin
0
1
100% 1.9%
To place the unit back into the Normal Operating mode:
1. With the scale on or off, toggle the Calibration Switch back to its original position. Turn the scale
back on without pressing any keys.
2. The display will go through a digit check, then settle into Normal Operating mode. All front panel
keys will now return to their normal mode of operation.