APPENDIX C: Displayed Error Codes

CODE

MODE

MEANING / POSSIBLE SOLUTION

Normal Operating Mode

A weight greater than 102% of the scale’s capacity has

been applied to the scale. Remove the weight from the

platter. Try re-calibrating the scale if this doesn’t solve the

problem. Otherwise, possible load cell damage due to

overloading.

_ _ _ _ _ _

Normal Operating Mode

Underrange condition. Platter has been removed. Re-install

platter. Try re-calibrating the scale if this doesn’t solve the

problem. Otherwise, possible load cell damage due to

shock loading.

– HI –

Warmup Mode

The scale has been turned on with a weight already present

on the platter. Remove the weight and then press the

ZERO

key if necessary. Try re-calibrating the scale if this doesn’t

solve the problem. Otherwise, possible load cell damage

due to overloading.

– LO –

Warmup Mode

The scale has been turned on without the platter installed.

Install the platter and then press the

ZERO

key if neces-

sary. Try re-calibrating the scale if this doesn’t solve the

problem. Otherwise, possible load cell damage due to

shock loading.

Err 0

Calibration Mode

Keyed-in weight value in Calibration Mode is less than 1.5

lb (0.7 kg). Use a larger test weight or re-adjust value.

Err 1

Calibration Mode

Keyed-in weight value in Calibration Mode is larger than

150 lb (70 kg). Use a smaller test weight or re-adjust value.

Err 2

Calibration Mode

Internal resolution is not high enough to process keyed-in

weight value in Calibration Mode. Verify test weight and

value.

Err 3

All Modes

Diagnostics check error - EEPROM Read

Err 4

All Modes

Diagnostics check error - EEPROM Write