Note, Diagram 4 – Arc Machines 415 WDR User Manual
Page 40

Model 415
Operation Manual
5.7
ARC VOLTAGE CONTROL (AVC) CALIBRATION
NOTE
A weld schedule is required for the following calibration. The first time a calibration is done it will be
necessary to CREATE a Calibration Weld Schedule. This section will outline all pertinent schedule
elements needed. (Pass 2)
1.
Insure that a Weld Head with AVC is connected per section III of the M415 Operation Manual.
NOTE
A weld schedule is required for the following calibration. The first time a calibration is done it will be
necessary to CREATE a Calibration Weld Schedule. This section will outline all pertinent schedule
elements needed. (Pass 2)
2.
Insure that a Weld Head with AVC is connected per section III of the M415 Operation Manual.
3.
The AVC Calibration is done on the AVC / OSC PCB (P/N 13D150313-01) located on the Servo
PCB mounting plate in labeled slot B 0/1. Locate the AVC adjustment trim pot R-40 on the
PCB (diagram 4). DO NOT ADJUST TRIM POTS R-80 AND R-81 (these are factory calibrated
for use with special application weld heads or fixtures not discussed in this section).
R-40
R-80 & R-81
DIAGRAM 4
4.
To calibrate the AVC the user will need to have a Weld Schedule in the Library for the M-415
and applicable Weld Head, (Pass 2).
5.
The AVC calibration schedule will be run in the WELD Mode and will contain the following:
Onelevel Functions
Prepurge
010
Torch Gas Rate
20
Upslope
01.0
Start Level
25
AVC Start Delay
1.0
Num Starts
1
Level
Advance
Manual
Postpurge
010
Stubout Mode
On
Downslope
05.0
Start Mode
Touch
AVC Stop Delay
1.0
AVC Resp
15**
Osc
Mode
OFF
**Note: For M-415 Software Revisions prior to Rev. 1.43, use an AVC Response of five (5) in
place of the pictured AVC Resp 15.
Document No. 740084
Revision N
35