Controller Features Level LN

2

Controllers

The controller is designed to maintain the LN

2

level in the

unit within a user-defined range. The LN

2

level will be

maintained between the low level sensor and the high level

sensor. When the liquid level reaches the low level sensor,

LN

2

will be added to the refrigerator until it reaches the high

level sensor.

Basic Operation

1.) Automatic Fill: The control will open the solenoid

valve automatically when the liquid level falls below the

Low Level Sensor. It will continue filling until the High

Level Sensor is covered by liquid.

Manual Fill: The Start Fill button can be pressed at

any time and the solenoid valve will open. If the liquid

level is between High Level Sensor and the Low Level

Sensor, the solenoid will stay open until the Stop Fill

button is pressed or until the liquid level covers the

High Level Sensor. If the liquid level is above the High

Level Sensor, the solenoid valve will stay open while

the user presses the Start Fill Button but will close

when the user releases the button.

Please Note: The maximum time that the

valve will stay open when the liquid level is

above the High Level sensor is one

minute. The user can open the valve again

by simply releasing and then pressing the

Start Fill button again.

Alarm Conditions: An alarm condition occurs when

a sensor problem develops or the supply tank runs

low on LN

2

. When an alarm condition does occur, the

appropriate light on the front panel flashes and an

audible alarm is activated.

Testing the front panel lights: To test all the lights on

the control except the Filling LED, press the Stop Fill &

Mute button and hold for 8 seconds.

Testing the Remote Alarm: To test the remote alarm,

press the Stop Fill & Mute button and hold for 13

seconds (5 additional seconds after testing the lights,

or unplug the power supply from the wall.)

Remote Alarm Jack: The remote alarm relay has a

set of “dry contacts” capable of carrying 5 amperes

current at 30 volts D.C. The relay is “normal” during

any alarm condition. The remote alarm is triggered 30

minutes after an error condition occurs. The remote

alarm will be reset when the error condition is cor-

rected. Pins 1 and 2 are closed in normal operating

condition while pins 2 and 3 are open in a remote

alarm condition. See Figure 8.

DESCRIPTION OF FRONT PANEL

Key Lock: This turns the control On/Off. Turning the Key to

the 3 o’clock position provides power to the control while

rotating the key 12 o’clock position turns the control off.

Start Fill: This button opens the solenoid valve and allows

LN

2

to flow into the freezer.

Stop Fill & Mute: This button closes the solenoid valve and

stops the flow of LN

2

into the freezer. This button also

silences the audible alarm.

Filling LED: Lights green to indicate that the solenoid valve

is open.

LN

2

Level LED: Lights red to indicate that the liquid level is

above the high level sensor. Lights green to indicate that

the liquid level is between the low level sensor and the high

level sensor. Lights yellow to indicate that the liquid level is

below the low level sensor.

Please note: The LED will not light if the high level

sensor is submerged in LN

2

while the low level

sensor is located in gas. The only time that this can

occur is if the sensors are installed backwards.

Sensor Fault LED: Lights red to indicate that a sensor fault

has occurred. A sensor fault can be either an open circuit or

a short circuit in the sensor assembly.

Low LN

2

Supply LED: Lights red to indicate that the LN

2

supply is low. This is triggered when the liquid level does

not reach the high level sensor within 1 hour of opening the

solenoid valve.

24VAC Power

Remote Alarm

24VAC Solenoid Valve

Sensor Assembly

Electrical Supply Connections For LS6000