5led indicators, 1 the power-led, 2 the laser off-led – Baumer BA DenexEdgeDetectorII User Manual
Page 10: 3 the output-led
User Manual DENEX Edge Detector II
10/20
Baumer Electric AG
Version 2013-02, V1.6
www.baumer.com
Frauenfeld, Switzerland
5
LED INDICATORS
There are four LED’s that are placed close to the connector.
Power (Green)
Laser off (Red)
Output (Orange)
Not used ( ellow)
Y
5.1 The POWER-LED
The green POWER-LED indicates that the sensor has power and should be operating.
For laser safety reasons never mount or adjust the sensor when Power-LED is on!
5.2 The LASER OFF-LED
The LASER OFF-LED is an indication that the laser is off. The laser will be shut off for the
following reasons:
•
The laser power has exceeded the allowed limits (electronic problem inside sensor)
•
The laser power has reached lower than acceptable level (laser is worn out, or
electronic problem)
If the laser levels are outside of the allowed limits, the sensor must be sent for repair.
5.3 The OUTPUT-LED
The OUTPUT-LED is active when the output stage is active (set). The LED will flash for
every output pulse.