Error Codes Table

Failure to reach speed in 30 seconds.
Confirm no obstruction in bowl area,
contact customer service.

Latch is not locked.
Confirm cover is fully seated in closed
position.

Rotor exceeded set speed.
Remove and inspect rotor, reinstall.

Cover is not closed.
Confirm cover is fully seated in closed
position.

Out of balance.
Confirm tubes are balanced in rotor and
centrifuge is on a level surface.


Locked Rotor
Contact customer service.
Confirm no obstruction in bowl area

Shield is not installed.
Confirm shield is installed and in the
locked position.


Continuously illuminated
Error / Service indicator.

Cycle Counter
50,000 spins have been reached, please
conduct periodic maintenance check (See
Section 4).

Accessories

Product No.

Description

SV02

Inserts for 3 mL tubes (10.25 x 64 mm) 4/pk.

SV05

Inserts for 2 mL tubes (10.25 x 47 mm or BD Microtainers) 4/pk.

SV06

Inserts for 7 mL tubes (16 x 75 mm) 8/pk.

SV07

Inserts for 10 mL tubes (16 x 100 mm) 8/pk. Included with centrifuge

SV08

Inserts for 5 mL BD tubes (13 x 75 mm) 8/pk.

SV09

Inserts for 7 mL tubes (13 x 100 mm) 8/pk. Included with centrifuge

SV10

Inserts for 5 mL Greiner tubes (13 x 75 mm) 8/pk.

SV11

Inserts for 4 mL BD RST tubes (13 x 100 mm) 8/pk

RTH8

8 place horizontal rotor. Included with centrifuge

3 min

5 min

3 min

5 min

10 min

10 min

10 min

3 min

3 min

5 min

5 min

10 min