3

Locked

Locked position for shield

Unlocked

Unlocked position for shield

Operator Controls

Start

The Start button initiates the selected cycle.

Stop/Open

The Stop/Open button interrupts the cycle and
releases the cover when the rotor has stopped.

3 Minute Cycle

The 3 min button selects a cycle of three minutes at
5100 rpm.

5 Minute Cycle

The 5 min button selects a cycle of five minutes at
5100 rpm.

10 Minute Cycle

The 10 min button selects a cycle of ten minutes at
4500 rpm.

Error/Service

The red indicator with an adjacent wrench symbol
flashes to signal an error condition or remains
continuously illuminated once the rotor reaches
50,000 cycles.

Error Indicators

StatSpin

®

Express 4 error conditions are signaled by a flashing red Error/Service indicator on the

control panel.

The cause of the error condition can be identified by the combination of green cycle time indicators
that are also flashing. See Error Codes Table.

To clear the error condition, correct the problem and press the Stop/Open button.
















3 min

5 min

10 min

Error/Service

light flashing