3 min 5 min 10 min – Iris Sample Processing StatSpin® Express 4 Horizontal Centrifuge User Manual
Page 7

3
Locked
Locked position for shield
Unlocked
Unlocked position for shield
Operator Controls
Start
The Start button initiates the selected cycle.
Stop/Open
The Stop/Open button interrupts the cycle and
releases the cover when the rotor has stopped.
3 Minute Cycle
The 3 min button selects a cycle of three minutes at
5100 rpm.
5 Minute Cycle
The 5 min button selects a cycle of five minutes at
5100 rpm.
10 Minute Cycle
The 10 min button selects a cycle of ten minutes at
4500 rpm.
Error/Service
The red indicator with an adjacent wrench symbol
flashes to signal an error condition or remains
continuously illuminated once the rotor reaches
50,000 cycles.
Error Indicators
StatSpin
®
Express 4 error conditions are signaled by a flashing red Error/Service indicator on the
control panel.
The cause of the error condition can be identified by the combination of green cycle time indicators
that are also flashing. See Error Codes Table.
To clear the error condition, correct the problem and press the Stop/Open button.
3 min
5 min
10 min
Error/Service
light flashing