Principle and Intended Use

For in-vitro diagnostic use to produce the rapid separation of whole blood contained in original collection
tubes.

The StatSpin

®

Express 4 is a high-speed horizontal bench top centrifuge used to rapidly separate blood

components in the original sample collection tubes. Do Not centrifuge substances with a density greater
than 1.2 kg/dm

3

.

The horizontal rotor included with the centrifuge accepts up to 8 primary 16x100mm primary blood tubes
with a maximum volume of 10 mL per tube. The operator selects a fixed spin time of 3, 5 or 10 minutes.
The rotation speed is

5100 rpm for the 3 and 5 minute cycles, producing a centrifugal force of

4000

xg at

a rotor radius of 13.8

cm. The 10 minute cycle generates 3100 xg at 4500 rpm. This setting provides

lower g-forces for procedures which require a traditional spin.

Symbols and Definitions

Product/Reference Number

Indicates the Iris Sample Processing

product/catalog number

Caution

Statement of caution/warning: Read

carefully.

Biohazard

Indicates potential biohazard

Temperature limitation

Indicates storage or operating

temperature restrictions

EC Representative

European Community Authorized

Representative

For in vitro diagnostic use

Indicates for use only as in-vitro diagnostic

device

Non-sterile

Indicates a non-sterile product

Date of manufacture

Indicates instrument manufacturing date

Serial Number

Indicates instrument serial number

Consult Instructions

Refers operator to the instruction manual

for additional information

European Community

Compliance to IVD Directive

Agency

Listed device with Intertek testing agency

Fuse

Rating: (located on serial number label,

replace with same value and type)

AC

Single Phase Alternating Current

IVD