beautypg.com

Appendix – Iris Sample Processing StatSpin® Express 4 Horizontal Centrifuge User Manual

Page 16

background image

12

Section

5

Appendix

Specifications

Product Number

SSH4

Model Number

M510

Spin Parameters

3 min: 180 seconds @ 5100 rpm / 4000 xg

5 min: 300 seconds @ 5100 rpm / 4000 xg

10 min: 600 seconds @ 4500 rpm / 3100 xg

Acceleration Time

≤ 30 seconds

Deceleration Time

≤ 30 seconds

Electrical

100

– 240 VAC, 50/60Hz, 2.5 Amps

Dimensions

Height: 8.0

“ (20.3 cm)

Width: 13.0

” (33.0 cm)

Depth: 16.0

” (40.6 cm)

Weight: 20 lbs. (9 Kg)

Environmental

For indoor use only

Ambient temperature 2ºC to 40ºC

Maximum 80% RH between 2ºC and 30ºC,
decreasing to 50% RH at 40ºC

Maximum altitude 2000 m

Main supply voltage fluctuations should not exceed ±10%

Transient over-voltages according to installation category II

Pollution degree 2

Specifications are subject to change without notice.

WARNING

– Do not operate or store the centrifuge or its rotor below the 2

o

C

minimum operating temperature. At extreme low temperature (below -18

o

C) the rotor

material is subject to cracking under impact stress (such as dropping onto a hard
surface).