To install the rotor:
1. Place the rotor back onto the cone-shaped adapter. You may

need to turn the rotor slightly to line it up properly.

2. The rotor should slide onto the rotor cone freely.
3. Once a proper fit has been achieved, replace the rotor nut and

turn it until it is hand-tight, (a tool may be required).

4. Replace the tube holders and verify that they are seated

properly, (as shown on page 4).

5. It is recommended that the initial setup procedures be performed

to ensure that the rotor has been installed correctly and that no

damage has been done to the centrifuge during either the rotor

installation or possible rotor chamber cleaning. See page 4 for

this procedure.