Drucker Diagnostics Model 642VFD Plus Centrifuge User Manual

Page 2

background image

Page 2

* The rotor and rotor accessories are rated for a rotation frequency of 4,000 RPM.

Safety:

Lid Safety Switch: The lid is secured to the top of the cabinet by a latching

knob and pawl system. When the knob is rotated clockwise, the pawl grips

the underside of the cabinet opening and prevents the lid from opening. A

mechanical stop positions the pawl and prevents it from rotating completely.

When rotated to the stop position, the pawl makes contact with a micro–switch

mounted underneath the cabinet top. The lid safety switch prevents the

centrifuge from operating while the lid is open. An indicator light on the front

of the machine will light up when the lid has been latched properly.

Lid Safety Plus Interlock System: In addition to the Lid Safety Switch, the

642VFD-Plus has a true “0 RPM” lid locking system. The lid safety interlock

system keeps the lid locked at all times, (even during power failure), and

requires that the rotor be at rest in order to unlock the lid. The centrifuge will

not allow entry into the rotor chamber unless the centrifuge has power and the

rotor is stopped. To open the lid, make sure that the centrifuge is plugged in

and, with the rotor stopped, press the ‘OPEN / STOP’ button.

Note: After the centrifuge has started spinning, it may be possible to rotate

the lid knob enough to cause the pawl to lose contact with the lid safety

switch. If this happens, the centrifuge motor may lose power, but the lid

will still remain locked. If the knob is accidentally moved and this situation

should occur, rotate the knob fully clockwise to its stop position and the

centrifuge will resume operation.

Circuit Breaker: The 642VFD-Plus is protected with a 4 Amp circuit breaker

located at the rear of the machine mounted to the base. Any electrical short

circuit will cause the breaker to cut power to the machine.

Emergency Rotor Chamber Entry:

In the event of power failure, it may be impossible

to unlock the lid by conventional means. In this

case, entry into the rotor chamber may be made

by removing the latch label and using a pen to

manually disengage the locking mechanism (see

photo). Pull the mechanism towards the control

panel and then unlatch and open the lid. If the

unit is damaged, contact your authorised dealer or

The Drucker Company.

Calibration and Earth Ground Testing:

It is recommended that the top speed, ground continuity and line leakage be

tested every two years for continued safe operation.
Contact The Drucker Company for further information or testing availability.

Page 11

×

6

Optional Accessories:
0.25” tube cushions, p/n 9150

1.0” tube cushions, p/n 1525

×

6

Model Description:

The 642VFD-Plus is a continuous-duty, electronically-controlled horizontal

centrifuge with a lid safety interlock system. The unit is controlled by two electronic

push–button timers that have been preset for ten (10) minutes 3,800RPM & five

(5) minutes 1,800 RPM, for precise spin times and ease of use. Samples can be

safely viewed through the transparent lid. Entry into the machine is restricted during

operation by the safety interlock system. The 642VFD Plus features a lighted

control panel that displays the status of the machine, easily viewable from

a distance. For warranty information, turn to page 12.

Intended Use

General purpose laboratory centrifuge for sample separation.

Supplied Equipment*:

The following items come standard with each 642VFD Plus Horizon

centrifuge:
1. Operator’s manual

2. One (1) six–place horizontal rotor

3. Six (6) 100mm tube holders

4. Six (6) 75mm tube holders

WARNING: For the safety of both

the operator and service personnel,

care should be taken when using

this centrifuge if handling sub-

stances that are known to be toxic,

radioactive or contaminated with

pathogenic microorganisms. When

Risk Group II materials are used,

(as identified in the World Health

Organization “Laboratory Bio–Safe-

ty Manual”), a Bio–Seal should be

employed. In the event that materi-

als of a higher risk group are being

used, more than one level of pro-

tection must be provided. The use

of flammable or explosive materials

as well as those materials which

have a vigorous chemical reaction

is prohibited.

Table of Contents

Model Description

pg. 2

Supplied Equipment

pg. 2

Features

pg. 3

Specifications

pg. 3

Setup Location and Procedure

pg. 4

Front Panel Controls

pg. 5

Operation

pg. 6

Rotor Removal and Installation

pg. 7

Tube Holder Configurations

pg. 7,8

Care and Maintenance

pg. 9

Troubleshooting

pg. 10

Safety, Emergency Lid Removal

pg. 11

Calibration and Ground Testing

pg. 11

Replacement Parts

pg. 12

Available Accessories

pg. 12

p/n

7786021

p/n

7713031

p/n

7713033