beautypg.com

Initial setup procedure – Drucker Diagnostics Model 642VFD Plus Centrifuge User Manual

Page 4

background image

Setup Location:
1. Unpack the centrifuge and verify that all of the supplied equipment is present.

2. Choose a setup location which meets the following criteria:

a) A bench top clearance height of 20” is required in order to open the lid.

b) The clearance envelope is the space around the centrifuge which is required for

safety. Choose a setup location which will allow for a clearance envelope of at

least 24”

×

24”, (with the centrifuge at the center). No person or hazardous material

shall be permitted in the clearance envelope during operation. The operator time

within the envelope shall be limited to the time necessary for loading, unloading

and centrifuge operation only.

c) Proper ventilation is necessary to prevent the overheating of samples as well as pre-

mature failure of the centrifuge. Choose an area which will allow unencumbered air flow.

d) The centrifuge is designed to secure to the operating surface by four suction feet.

No adjustment is necessary for leveling the centrifuge, however, the surface should

be flat and level.

e)

Be sure the outlet is always within reach as the line cord is the means of

emergency

disconnection!

Initial Setup Procedure:

If any problems are found during the initial setup procedure, refer to the trouble-

shooting section on page 10.
1. Plug the female end of the line cord into the rear of the centrifuge. Plug the male end

into an approved electrical outlet. For electrical safety, the unit must always be

properly grounded.

2. For safety purposes, the locking system is always activated. To deactivate the system,

(in order to insert or retrieve samples), press the ‘OPEN / STOP’ button on the control

panel. The ‘UNLOCKED’ indicator light should illuminate. If it does not, refer to page 10 on

troubleshooting. The lid will be unlocked for 15 seconds after pushing the ‘OPEN / STOP’ button.

3. Turn the latch counterclockwise and open the lid.

4. Spin the rotor by hand; check for free and level rotation. If the rotor does not spin

freely, refer to the section on troubleshooting.

5. Place the six test tube holders inside the rotor (as shown

to the right), and verify that they are seated properly.

6. Close the lid. Rotate the lid knob clockwise to its complete

stop position. The ‘LATCH’ indicator light should be

illuminated. If it is not, make sure that the lid is latched

properly. The centrifuge will not run unless the lid is

latched and the ‘LATCH’ indicator light is illuminated.

7. Turn the centrifuge on by pushing the desired ‘START’

button; either PRP or PRFM.

8. The appropriate ‘RUN’ indicator light will illuminate.

9. The test tube holders will slide up into the horizontal position and the unit will

accelerate to full speed.

10. Listen to the sound of the centrifuge. A smooth whirring sound should be heard.

If there are any loud or unusual sounds, stop the centrifuge by pressing the ‘OPEN/

STOP’ button immediately and refer to the section on troubleshooting.

11. Push the ‘OPEN / STOP’ button. The ‘RUN’ indicator light should go out and the motor

should slow to a stop.

12. The lid should remain locked until the rotor has nearly stopped. If the machine unlocks

prematurely, contact The Drucker Company for assistance. Once the rotor has stopped,

the interlock system will become disengaged for sixty (60) seconds. The ‘UNLOCKED’

indicator light will illuminate during this time.

13. To gain entry into the centrifuge after this period has ended, simply press the

‘OPEN / STOP’ button. The lid will unlock for fifteen (15) additional seconds.

After the centrifuge has passed this procedure it is ready for operation.

Page 4

(cont.)

Care and Preventative Maintenance:
With proper care and maintenance your 642VFD-PLUS centrifuge will provide

years of laboratory service. For proper care, the following steps should be taken:

1.

Provide Adequate Ventilation: For cooling purposes, the 642VFD-PLUS draws

in ambient air through the air intake cover on the top of the lid and exhausts

this air in the rear of the base. The centrifuge should be placed on a hard smooth

surface for good air circulation.

2.

Always Spin Balanced Loads: Make certain that you are always spinning a

balanced load. The 642VFD-PLUS has a unique counter balanced motor

mounting design which, along with it’s rubber suction feet, produces excellent

vibration dampening. However, out–of–balance loads may break glass test

tubes and may produce unsatisfactory separation results. Proper load balancing

will improve sample separation and extend the life of the centrifuge. See page

6 for information on balancing the load.

3.

Keep the Tube Carriers / Holders Clean: NOTE: Always follow the safety

guidelines of your laboratory to properly clean up and/or dispose of materials

in the event that a substance known to be potentially toxic, radioactive or

contaminated with a pathogenic microorganism is spilt in or on the centrifuge.

Small glass fragments left in the tube holder after a tube breakage may adhere

to the next test tube inserted in that holder. When this tube is handled, these

fragments may puncture protective gloves and lacerate the operator’s fingers

or hand. Remaining fragments may provide stress points on subsequent tubes

and result in additional breakage. If a tube breakage occurs, carefully remove the

tube holder. Properly dispose of the sample and tube fragments and thoroughly

clean both the inside and outside of the tube holder. Insert a new tube cushion

(if necessary) and replace the tube holder in the rotor.

4.

Motor and Electrical Maintenance: The 642VFD-PLUS uses a brushless

permanent split capacitor AC motor. It should not need servicing for the life

of the centrifuge. The electrical components are selected for high reliability and

should not need service.

5.

Keep the Rotor Chamber Clean: Every six months, or whenever there is a tube

breakage, (refer to the note in #3), it may be necessary to remove the rotor and

clean the rotor chamber. Follow the instructions on page 7 to remove and reinstall

the rotor.

CAUTION: Once the lid has been opened, unplug the line cord from the electrical

outlet to eliminate the risk of electric shock during cleaning.

The rotor chamber, rotor and accessories shall be thoroughly cleaned using

either isopropyl alcohol, soap and water, or bleach. The use of Fully / Partially

Halogenated Hydrocarbons, Ketones, Esters and all other chemicals not

prescribed by the manufacturer may cause damage to the rotor and tube holders

and shall not be used.

6.

Tube Holder Replacement: It is recommended that tube holders be replaced

after 24 months of use.

Page 9

Horizontal Rotor Shown