Thoroughly inspect instruments for effective cleaning (free
from any visible dirt and residue). Also inspect the
instruments for any damaged, worn, corroded, pitted/porous
conditions and ensure proper function. Contact Miltex if
damage is observed and if repair is necessary. Apply
non-silicone instrument lubricant (such as Miltex REF 3-700
Instrument Spray Lube) to the joints and jaws of the instrument
to ensure easy movement. Caution: Do not oil electrical
connections on instruments.
Ultrasonic Cleaners:
Ultrasonic Cleaners can also be used for cleaning and removal
of contamination and debris. Follow the Ultrasonic Cleaner
manufacturer’s recommendations for cleaning agents, fill
levels, temperature, processing time, etc. Make sure that
functional parts do not touch other instruments and all jointed
instruments are processed in an open position. We
recommend pre-flushing instruments via the equipped
LUER-LOCK connection which should be opened and force flushed
using the appropriate equipment prior to immersion in the
ultrasonic bath. After cleaning and disinfection, the
instruments should be thoroughly rinsed, both on the exterior
and interior sheath, with distilled water. Immediately dry
instruments to prevent corrosion. Drying can be performed
manually, but compressed air pistol is recommended for more
delicate instruments and is more effective for overall drying.
Thoroughly inspect instruments for effective cleaning (free
from any visible dirt and residue). Also inspect the
instruments for any damaged, worn, corroded, pitted/porous
conditions and ensure proper function. Contact Miltex if
damage is observed and if repair is necessary. Apply
non-silicone instrument lubricant (such as Miltex REF 3-700
Instrument Spray Lube) to the joints and jaws of the instrument
to ensure easy movement. Caution: Do not oil electrical
connections on instruments.
Warning: Ultrasonic cleaning should never be used for
endoscopes and instruments with tungsten carbide inserts.
2. STERILIZATION
Steam sterilization should be carried out by steam under
pressure in the complete absence of air in an autoclave set at
134º C as follows:
• Temperature 134°C, Sterilizing time = 5 min.
Instrument jaws and ratchets must be in open position to avoid
damage during the sterilization heating and cooling process.
Always follow the autoclave manufacturer instructions and
recommendations. For autoclaving, place the instruments in a
suitable container or sterilization tray. Cutting edges and tips
of instruments should always be protected with a soft cloth.
After autoclaving allow all components cool down to room
temperature. Sudden changes in temperature may cause
damage to the instruments.
3. GAS STERILIZATION
Place the pre-cleaned instrument in a transparent aseptic
package. It is recommended to cover tips and cutting edges of
instruments to avoid perforation of the packaging. Follow the
instruction manual of the manufacturer of the gas sterilization
equipment and observe domestic health care regulations.
Allow sufficient time for ventilation to remove toxic residues.
4. SERVICE
Contact the Miltex Repair department at toll free phone
number 1-866-854-8300. Miltex does not supply its
mechanical parts to any outside company, therefore only Miltex
or its subsidiaries are able to repair Miltex
®
brand laparoscopic
instruments with original parts and to guarantee the original
technical specification and safety of the product.
Additionally, this warranty is void for Miltex
®
instruments
and products serviced by any person or facility other than
the Miltex Instrument Repair Services.
5. WARRANTY
Miltex
®
laparoscopic surgical hand instruments have a 1-year
limited warranty against defects in workmanship and material
which excludes wear to blades and jaws. Any Miltex
laparoscopic instruments that prove defective in
workmanship or material will either be repaired or replaced
at our discretion, without charge. Our liability under this
warranty shall be limited to repair or replacement of defective
merchandise. This warranty is a limited warranty and is the
only warranty made by Miltex. Miltex makes no other warranty,
expressed or implied, including any warranty of merchantability
and fitness for a particular purpose.
INSTRUCTIONS FOR CARE & CLEANING OF
MILTEX
®
ENDOSCOPES AND LAPAROSCOPIC
SURGICAL HAND INSTRUMENTS
Instructions for Use/English