Humboldt HM-1327 Humboldt Marshall/TSR Loader User Manual
Page 21

18
Press
(SELECT INPUT) to scroll through and choose the channel you
wish to calibrate. Once you have chosen the desired channel you wish to 
calibrate press 
(CLEAR CALIBRATION).
WARNING: IF YOU CLEAR A CALIBRATION IT IS LOST FOREVER! 
 
After	the	calibration	has	been	cleared,	Press	
(CALIBRATE CHANNEL)
You should see this screen:
Press
(SET GAIN). With the instrument you wish to calibrate
mounted to a calibration frame attached to the HM-1327, record the sensor 
min value with no load on the instrument (X divisions). 
Next,	apply	the	full	engineering	load	value	to	instrument	on	the	calibration	
frame and again, record the sensor min reading (X divisions). Subtract the 
two values. This value should be a number larger than the sensor maximum 
rating.
If	not,	you	need	to	increase	the	gain	to	compensate	for	the	difference.	Press
(SET GAIN) and repeat until the difference between the two values is
more than the maximum sensor rating.
EXAMPLE: 
If you are calibrating a 10,000 lb. load cell, you need a difference value of 
at least 10,000 divisions. If you register a min value reading of 450 divisions 
with no load on the sensor and your min is 8450 divisions at full load on 
the instrument, your difference value will be 8000 divisions. (8450 – 450 = 
8000). To increase the gain press 
(SET GAIN) until you have more
than 10,000 divisions.
Once the gain has been set, the zero point needs to be set. To set the zero 
point, apply the 0% force to the sensor and press 
(SET SENSOR MIN).
Now apply the 100% force to the sensor and press
(SET SENSOR
MAX).
To return to any prior screen you can always press
On the next page is an Instrumentation calibration chart listing max 
divisions, as well as typical gain factors needed in calibrating each 
instrument type.
INPUT CALIBRATION 
* SET GAIN 
1
* SET SENSOR MIN
123
* SET SENSOR MAX 12345