Bio-Rad Foresight™ Chromatography Columns, Prepacked User Manual
Page 19

Technical Support: 1-800-424-6723 • www.bio-rad.com 15
Section 5
Troubleshooting Guide
Fault
Possible cause
Action
Foresight
™
RoboColumn 
units are 
clogged
•
The sample is too 
viscous
•
There is too much 
cell debris in the 
sample
•
The column has 
been regenerated 
and used for too 
many cycles
•
Increase dilution of 
the cell isolate before 
lysis, or dilute after 
the lysation
•
Centrifuge and/or 
filter the sample if 
unclarified sample 
has been used
•
Reduce the number 
of cycles
Foresight 
RoboColumn 
units are leaking
•
Columns are 
clogged
•
The O-ring seal is 
damaged
•
The insertion depth 
of the robotic needle 
is incorrect
•
See above
•
Check the taught 
X and Y coordinates 
for RoboColumn 
labware for precision
•
Check the insertion 
depth (16 mm 
from the top of 
the column) of the 
robotic needle
Foresight 
RoboColumn 
units are lifted 
up from the 
96-well array 
plate during the 
pull-out of the 
robotic needles
Columns are not securely 
seated into the 96-well 
array plate
Make sure that all 
columns are securely 
seated into the 
96-well array plate by 
firmly pressing down 
before starting the 
experiment
Low protein 
recovery in the 
elution fraction(s)
The residence time for 
sample loading is too low
Decrease the flow rate or 
increase residence time 
for sample loading