Bio-Rad ProteinChip Qualification and Calibration Kits User Manual
Page 5

Table 1. Summary of procedures and tests.
C
Co
on
nssu
um
maab
bllee
E
Essttiim
maatteed
d T
Tiim
mee
T
Tiittllee
U
Usseed
d
R
Reeq
qu
uiirreed
d
D
Deessccrriip
pttiio
on
n
IIn
niittiiaall S
Seettu
up
p
File and protocol
20 min
Complete this procedure
setup
the first time you use the
ProteinChip system check
kit. This procedure creates
the file structure necessary
for storing the protocols
and creates the routines for
high voltage conditioning.
M
Maaiin
ntteen
naan
nccee
High voltage
N/A
1–2 hr
HV conditioning helps to
P
Prro
occeed
du
urree 1
1
(HV) depending
on
decontaminate the surfaces
conditioning
instrument
in the instrument.
M
Maaiin
ntteen
naan
nccee
Detector
ProteinChip
45 min to 4 hr
This procedure uses the
P
Prro
occeed
du
urree 2
2
calibration
detector
depending on
ProteinChip detector
calibration
instrument
calibration array to adjust
array
the detector voltage. These
adjustments are based on a
rolling average that stabilizes
the gain, improving spectral
reproducibility over the
lifetime of the detector.
T
Teesstt 1
1
Detector
ProteinChip
30 min to 1 hr
This test uses the
sensitivity
detector ProteinChip
detector
qualification
calibration array to
array
measure the
signal-to-noise ratio
(S/N) of immunoglobubulin
(IgG) at two different
concentrations (10 fmol
and 140 fmol).
Measurements are
compared to a specification,
and a pass/fail disposition is
obtained.
2
10010681A.qxp 8/17/2007 11:22 AM Page 2