method parameters shown in Table 3. The Shutdown Method
included in the BioFocus software can be used as a template to
quickly program the dsDNA Shutdown Method.

Table 3. dsDNA Shutdown Method

ID: END

Description: terminates autosequence

Prep 1: Pre-Inject from [WATER

] to [Waste

]

180 sec

Prep 2: Pre-Inject from [NITROGEN

] to [Waste

]

300 sec

No Injection

Polarity:

Negative -> Positive

Run Voltage:

0.00 kV

Current Limit:

0.30 mA

Inlet buffer:

[WATER

]

Outlet buffer:

[WATER

]

Cartridge
Temperature:

20 degrees Celsius

Run Time:

1.00 min

Detector:

LIF

Channel: 1

Laser(s) turned off at the end of the run group

9

Table 2. dsDNA Analysis Method for 24 cm Capillary

Section 4
System Shutdown Method

Drying of the buffer at the ends of the capillary may cause

the capillary to become plugged. For short term storage (24 hours),
leave the capillary filled with the CE dsDNA run buffer and place
the capillary ends in test tubes filled with deionized water at 4 °C.
For long term storage, the capillary should be flushed thoroughly
with water and then purged with nitrogen. This may be programmed
to run automatically at the end of an automation sequence using the

8

Set the carousel temperature to 20 °C