Bio-Rad ReadyPrep Protein Extraction Kit (Soluble/Insoluble) User Manual
Page 14
Section 6
Appendix
2-D Rehydration/Sample Buffer Volume
Before IEF and 2D gel electrophoresis, the protein sample
may need to be diluted to achieve the desired protein load
for the chosen stain. To best determine the volume of
diluent to use, consider the IPG strip length, the pH
gradient of the IPG strip, and the staining or detection
method. To assist with these calculations, the table that
follows indicates appropriate volumes of 2-D
rehydration/sample buffer needed to rehydrate IPG strips of
specific lengths and the approximate amounts of protein
required for detection using silver stain or Coomassie Blue
G-250 stain.
IPG strip length
7 cm
11 cm
17 cm
18 cm
24 cm
Rehydration
125 µl
185 µl
300 µl
315 µl
410 µl
volume per strip
Protein load–
5–20 µg
20–50 µg
50–80 µg
50–80 µg
80–150 µg
Silver stain
Protein load–
50–100 µg
100–200 µg
200–400 µg
200–400 µg
400–800 µg
Coomassie G-250
12