Problem
Possible Cause
Recommended Solution
Low yield
Insufficient amount of
Scale up the amount of
(continued)
PureZOL used for
PureZOL used according to
sample lysis and
sample size. Sample volume
homogenization
should not exceed 10% of
the PureZOL reagent used
for homogenization
Lysate was not
Make sure to incubate the
incubated at room
lysate after the disruption
temperature for
step for 5 minutes at room
5 minutes after the
temperature to allow
disruption step (see
complete dissociation of
step 4 in protocol)
nucleoprotein complexes
RNA is
RNase contamination
Treat all user-made solutions
degraded
of solutions supplied
with DEPC before use (see
by the user
section 3, Maintaining an
RNase-Free Environment)
RNase contamination
See section 3, Maintaining
of plasticware and
an RNase-Free Environment
work station
Frozen tissue samples
Add PureZOL directly to
were allowed to thaw
frozen samples before they
or sit at room
thaw. Do not let starting
temperature
samples sit at room
temperature
14
10001481A.qxp 3/7/2005 7:05 AM Page 14
- S3e™ Cell Sorter (36 pages)
- S3e™ Cell Sorter (132 pages)
- iMark Microplate Absorbance Reader (112 pages)
- xMark™ Microplate Absorbance Spectrophotometer (20 pages)
- Microplate Manager Software (138 pages)
- iMark Microplate Absorbance Reader (48 pages)
- Model 680 Microplate Reader Accessories (150 pages)
- Model 550, Ultramark, and Benchmark Microplate Reader Accessories (31 pages)
- Model 550, Ultramark, and Benchmark Microplate Reader Accessories (39 pages)
- Model 550, Ultramark, and Benchmark Microplate Reader Accessories (15 pages)
- Model 680 Microplate Reader Accessories (68 pages)
- Mouse Typer Isotyping Kit (14 pages)
- ZOE™ Fluorescent Cell Imager (36 pages)
- TC20™ Automated Cell Counter (34 pages)
- Design Your NGC™ System (306 pages)
- Design Your NGC™ System (300 pages)
- ChromLab™ Software (178 pages)
- BioLogic DuoFlow 10 System V9-8 (4 pages)
- BioLogic DuoFlow 10 System HR Gradient Pump Maintenance Kit (4 pages)
- BioLogic DuoFlow 10 System DuoFlow Chromatography System Version 4.0 Software New Features (7 pages)
- BioLogic DuoFlow 10 System DuoFlow Chromatography System Starter Kit (40 pages)
- BioLogic DuoFlow 10 System DuoFlow Chromatography System 5.0 Software New Features (6 pages)
- BioLogic DuoFlow 10 System pH Monitor (4 pages)
- BioLogic DuoFlow 10 System Chromatography System (239 pages)
- BioLogic Mixers (2 pages)
- BioLogic QuadTec™ Detector and Components (58 pages)
- BioLogic Sample Loading Loops (20 pages)
- LP Data View Software for BioLogic LP System (5 pages)
- BioLogic LP System (106 pages)
- LP Data View Software for BioLogic LP System (4 pages)
- BioLogic LP System (28 pages)
- BioLogic LP System (5 pages)
- BioLogic LP System (1 page)
- Model EP-1 Econo Pump (1 page)
- Model EP-1 Econo Pump (23 pages)
- Econo Gradient Pump Kits (19 pages)
- Econo Gradient Pump Kits (51 pages)
- Model EP-1 Econo Pump (27 pages)
- Profinia™ Protein Purification Instrument (4 pages)
- Profinia™ Protein Purification Instrument (2 pages)
- Profinia™ Protein Purification Instrument (224 pages)
- Profinia Protein Purification System Accessories (1 page)
- Profinia Protein Purification System Kits (24 pages)
- Profinia Protein Purification System Kits (24 pages)
- Profinia Protein Purification System Buffers and Reagents (4 pages)