Bio-Rad PureZOL™ RNA Isolation Reagent User Manual
Page 14
![background image](/manuals/600459/14/background.png)
Section 6
Troubleshooting
Problem
Possible Cause
Recommended Solution
Incomplete
Lysate was not mixed
Once chloroform is added,
separation of
properly after adding
mix tubes vigorously by
phases after
chloroform (See step 3
shaking for 15 seconds. Do
centrifugation
in protocol)
not vortex. Let the lysate
incubate for 5 minutes at
room temperature, then mix
again before centrifuging
Lysate was not
Make sure that the
centrifuged at the right
centrifugation is performed
temperature
at 4°C following the addition
of chloroform in order to
achieve complete separation
of the phases
Incorrect amount of
For every 1 ml of PureZOL
chloroform was added
used, add 0.2 ml of
chloroform
Low yield
Incomplete disruption
Increase the duration or
and homogenization
intensity of sample
of sample
disrup tion.Scale up the
amount of PureZOL used
based on sample size
Over dried RNA pellet
Do not use a Speed Vac to
dry the pellet. Dry pellet
briefly at room temperature
for 5 minutes. Add DEPC-
treated water to the tube
and pipet up and down
several times to resuspend
12
10001481A.qxp 3/7/2005 7:05 AM Page 12
- S3e™ Cell Sorter (36 pages)
- S3e™ Cell Sorter (132 pages)
- xMark™ Microplate Absorbance Spectrophotometer (20 pages)
- iMark Microplate Absorbance Reader (112 pages)
- Microplate Manager Software (138 pages)
- iMark Microplate Absorbance Reader (48 pages)
- Model 550, Ultramark, and Benchmark Microplate Reader Accessories (39 pages)
- Model 550, Ultramark, and Benchmark Microplate Reader Accessories (15 pages)
- Model 680 Microplate Reader Accessories (150 pages)
- Model 550, Ultramark, and Benchmark Microplate Reader Accessories (31 pages)
- Model 680 Microplate Reader Accessories (68 pages)
- Mouse Typer Isotyping Kit (14 pages)
- ZOE™ Fluorescent Cell Imager (36 pages)
- TC20™ Automated Cell Counter (34 pages)
- Design Your NGC™ System (306 pages)
- Design Your NGC™ System (300 pages)
- ChromLab™ Software (178 pages)
- BioLogic DuoFlow 10 System V9-8 (4 pages)
- BioLogic DuoFlow 10 System HR Gradient Pump Maintenance Kit (4 pages)
- BioLogic DuoFlow 10 System DuoFlow Chromatography System Version 4.0 Software New Features (7 pages)
- BioLogic DuoFlow 10 System DuoFlow Chromatography System Starter Kit (40 pages)
- BioLogic DuoFlow 10 System DuoFlow Chromatography System 5.0 Software New Features (6 pages)
- BioLogic DuoFlow 10 System pH Monitor (4 pages)
- BioLogic DuoFlow 10 System Chromatography System (239 pages)
- BioLogic Mixers (2 pages)
- BioLogic QuadTec™ Detector and Components (58 pages)
- BioLogic Sample Loading Loops (20 pages)
- BioLogic LP System (5 pages)
- BioLogic LP System (1 page)
- LP Data View Software for BioLogic LP System (5 pages)
- BioLogic LP System (106 pages)
- LP Data View Software for BioLogic LP System (4 pages)
- BioLogic LP System (28 pages)
- Model EP-1 Econo Pump (1 page)
- Model EP-1 Econo Pump (23 pages)
- Econo Gradient Pump Kits (19 pages)
- Econo Gradient Pump Kits (51 pages)
- Model EP-1 Econo Pump (27 pages)
- Profinia™ Protein Purification Instrument (4 pages)
- Profinia™ Protein Purification Instrument (2 pages)
- Profinia™ Protein Purification Instrument (224 pages)
- Profinia Protein Purification System Accessories (1 page)
- Profinia Protein Purification System Kits (24 pages)
- Profinia Protein Purification System Kits (24 pages)
- Profinia Protein Purification System Buffers and Reagents (4 pages)