Linear flow rate, cm/h operating pr essur e, psi, Operating pr essur e, bar – Bio-Rad Media Sampler Pack User Manual
Page 5
3
400
300
200
100
0
0
10
20
30
40
Linear flow rate, cm/h
Operating pr
essur
e, PSI
0.0
1.0
2.0
Operating pr
essur
e, bar
Fig. 1. Linear flow rate vs. operating pressure. Data for Macro-Prep High Q support was generated
on a 14 cm (ID) X 17.4 cm (H) Amicon Moduline glass column with 1/4” plumbing. A Cole-Parmer Masterflex
pump, size 18 neoprene tubing, Easy-Load pumphead was used. The eluent was water.
Table 1. Properties of Macro-Prep Ion Exchange Supports.
Macro-Prep High Q Macro-Prep DEAE
Macro-Prep High S Macro-Prep CM
Support
Support
Support
Support
Type of ion exchanger
Strong anion
Weak anion
Strong cation
Weak cation
Functional ligand
–N
+
(CH
3
)
3
–N
+
(C
2
H
5
)
2
–SO
3
–
–COO
–
Ionic capacity
400 ± 75 µeq/ml
175 ± 75 µeq/ml
160 ± 40 µeq/ml
210 ± 40 µeq/ml
Dynamic binding capacity
≥37 mg BSA/ml
≥30 mg BSA/ml
≥49 mg IgG*/ml ≥25 mg Hemoglobin/ml
Counterion
Cl
–
Cl
–
Na
+
Na
+
Nominal particle size†
50 µm
50 µm
50 µm
50 µm
Nominal pore size
1,000 Е
1,000 Е
1,000 Е
1,000 Е
Recommended maximum
3,000 cm/hr
3,000 cm/hr
3,000 cm/hr
3,000 cm/hr
linear flow rate
Chemical stability
1% SDS, 24 hr
Yes
Yes
Yes
Yes
6 M Guanidine-HCl, 24 hr
Yes
Yes
Yes
Yes
pH stability**
1–10
1–10
1–12
1–12
Volume changes
pH 4–10
<1%
<1%
<3%
<1%
0.1–1.0 M NaCl
<5%
<5%
<9%
<4%
Autoclavable (121°C, 30 min)
Yes
††
Yes
††
Yes
Yes
Antimicrobial agent
20% ethanol
20% ethanol
20% ethanol
20% ethanol
* Human IgG
**Based on leachables, as determined by TOC (see Table 2)
†
Based on underivatized beads
††
Do not autoclave in OH– form
LIT271D.qxd 5/23/2005 12:56 PM Page 3