Bio-Rad LP Data View Software for BioLogic LP System User Manual
Page 5
Saving a Log file as a Data file
If a Log file is saved as a Data file
To prevent the problem, enter a different
’ .
without entering a different name for
the tile, the original. Log file remains
name when saving a Log file as a Data
file.
but the Data file is not saved. Data
View also incorrectly presents a
warning about “replacing an existing
file”. When this problem occurs, no
data is lost.
Setting axes
Viewing multiple open
applications
If multiple entries are made in the Axes
the ‘Set” button as appropriate
Settings Text Boxes for different axes
for each axis.
before clicking on the “Set!’ button or
pressing the “Enter” key, you must
click on each “Set” button for all
entries to take effect. The “Enter” key
will only activate the “Set” button for
the axis Text Box which has the focus.
When data recording is started
The user should click on the program
automatically as the
they were using previously or use
System starts a run, Data View takes
to send the focus back to the
the computer focus away from other
desired program.
applications and does not return it even
if
Data View is minimized.
Scroll bars
While recording -data and using
The solution is to use the Full View
Window mode for the Time Axis, the
setting for the Time Axis to see all data
scroll bar is not active.
and then use the mouse again to “zoom
in” on a new region.
Viewing runs of less than one
minute duration
Data View does not use the scroll
The Full View setting will show all
bar correctly when using the Window
data and the mouse can be used to
setting for the Time Axis when less
rubberband the data if necessary.
than 1 minute of data is present.
Printing in Landscape format
The program does not print in
No work-around for this release.
Landscape mode in Windows NT.