beautypg.com

Plan plate layout – Bio-Rad Bio-Plex Pro™ TGF-β Assays User Manual

Page 11

background image

9

1.

Plan Plate Layout

Prior to running the assay, determine the total number of wells in the
experiment using the Plate Layout Template on page 36 or the Plate
Formatting tab in Bio-Plex Manager

software. A suggested plate layout is

shown in Figure 2, with all conditions in duplicate.

1. Assign standards to columns 1 and 2, with the highest
concentration in row A and the lowest concentration in row H.

2. Assign the blank to wells A3 and A4. The blank should consist of your

chosen standard diluent and be processed in the same manner as
sample and standard wells. Note that Bio-Plex Manager automatically
subtracts the blank (B) MFI value from all other assay wells.

3. User-defined controls are assigned to wells in columns 3 and 4.

4. The remainder of the plate is available for samples.

5. Once the total number of wells is known, calculate the required

volumes of beads, detection antibody, and streptavidin-PE needed.
Use Tables 7–8, 10–11, and 12, respectively, or the Calculation
Worksheet on pages 37–38.

Legend

S Standard

B Blank

X Samples

C Controls

Fig. 2. Suggested plate layout. For detailed instructions on
plate formatting in Bio-Plex Manager, see section 8, Read Plate.