1 overview – Bio-Rad PowerPac™ HC High-Current Power Supply User Manual
Page 7
Section 1
Introduction
1.1 Overview
The PowerPac HC power supply is designed to provide constant voltage, current or
power for electrophoresis applications. The power supply operates at the value
specified for the constant parameter, with limits for the other parameters. To prevent
damage to the electrophoresis cell, the PowerPac HC provides automatic
crossover to constant current, constant voltage, or constant power, depending on
which set limit is first reached. If the set limit of the non-constant parameter is
reached, the power supply will switch, making the limiting non-constant parameter
the new constant.
Output specifications:
Voltage:
Adjustable from 5 to 250 volts (V), in 1 V increments
Current :
Adjustable from 0.01 to 3.00 amperes (A), in 0.01 A increments
Power:
Adjustable from 0.05 to 300 watts (W) in 1 W increments
Output jacks:
Four sets of output jacks are provided to facilitate connection of
up to 4 identical electrophoresis cells simultaneously
Fig. 1. Front View.
Fig. 2. Rear View.
setup
stop/home
abc
1
2
def
3
jkl
ghi
4
5
mno
6
tuv
pqrs
7
EDIT
8
0
CE
wxyz
9
OUTPUT JACKS
FRONT PANEL
1
- TransFectin™ Lipid Reagent (2 pages)
- Gene Pulser MXcell™ Electroporation System (19 pages)
- Gene Pulser MXcell™ Electroporation System (66 pages)
- Gene Pulser Xcell™ Electroporation Systems (83 pages)
- Gene Pulser® Electroporation Buffer (2 pages)
- MicroPulser™ Electroporator (31 pages)
- Helios® Gene Gun System (52 pages)
- PDS-1000 / He™ and Hepta™ Systems (51 pages)
- TGX™ FastCast™ Acrylamide Solutions (2 pages)
- Criterion™ TGX Stain-Free™ Precast Gels (16 pages)
- Criterion Stain Free™ Tris-HCl Gels (96 pages)
- Image Lab™ Software (212 pages)
- Gel Doc™ EZ System (22 pages)
- Mini-PROTEAN® TGX™ Precast Gels (52 pages)
- ChemiDoc™ MP System (8 pages)
- Image Lab™ Software (236 pages)
- ChemiDoc™ XRS+ System (42 pages)
- ChemiDoc™ XRS+ System (4 pages)
- ChemiDoc™ XRS+ System (50 pages)
- GS-800™ Calibrated Densitometer (444 pages)
- Criterion™ TGX™ Precast Gels (60 pages)
- Criterion™ Cell (13 pages)
- Image Lab™ Software (260 pages)
- Criterion™ XT Bis-Tris Precast Gels (26 pages)
- 2-D Electrophoresis Workflow (14 pages)
- EXQuest Spot Cutter (81 pages)
- Fluorescent Protein Stains (27 pages)
- 2-D Electrophoresis Workflow (22 pages)
- 2-D Electrophoresis Workflow (20 pages)
- PROTEAN® i12™ IEF System (4 pages)
- ReadyPrep™ 2-D Starter Kit (28 pages)
- Ready Gel® Zymogram Precast Gels (46 pages)
- Mini-PROTEAN 2-D Electrophoresis Cell (44 pages)
- Precision Plus Protein™ Prestained Standards (3 pages)
- Precision Plus Protein™ Unstained Standards (16 pages)
- Prestained SDS-PAGE Standards (3 pages)
- Silver Stains (20 pages)
- Unstained SDS-PAGE Standards (3 pages)
- Biotinylated Standards (3 pages)
- Biotinylated Standards (11 pages)
- IEF and 2-D Standards (3 pages)
- Mini-PROTEAN 2-D Electrophoresis Cell (20 pages)
- Mini-PROTEAN® Tetra Handcast Systems (10 pages)
- Mini-PROTEAN 3 Multi-Casting Chamber (10 pages)
- PROTEAN® Plus Hinged Spacer Plates and Combs (21 pages)