Bio-Rad ReadyPrep™ 2-D Starter Kit User Manual
Page 25

23
4. - The first line displays the step number, slope and voltage
- Select HRS:MIN (default)
- Enter time of 20 min.
- Press NEXT
S 01
250 V
HRS:MIN
vhours
>
ENTER TIME:
00:20
>
NEXT
>
5. - Select ADD (default)
- Enter Step 2
- Press NEXT
ADD inset delete >
ENTER STEP #
2 >
NEXT
>
METHOD COMPLETE
>
6. - Enter step 2 final voltage
7 cm: 4,000 V; 11 cm: 8,000 V; 17 cm: 10,000 V)
- Select LINEAR ramp. Use key next to arrow, >, to move
the brackets, [ ] ,around the selection
- Press NEXT
S 02
xxxx V
>
SELECT VOLTAGE SLOPE
LINEAR
[
д
д]
>
NEXT
>
7. - The first line displays the step number, slope and voltage
- Select HRS:MIN (default)
- Enter time
(7 cm: 2:00; 11 cm: 2:30; 17 cm 2:30)
S 02
xxxx V
>
HRS:MIN
vhours
>
ENTER TIME:
xx:xx >
NEXT
>
8. - Select ADD (default)
- Enter step 3
- Press NEXT
ADD Inset delete >
ENTER STEP #
3 >
NEXT
>
METHOD COMPLETE
>
9. - Enter step 3 final voltage
(7 cm: 4,000 V; 11 cm: 8,000 V; 17 cm: 10,000 V)
- Select RAPID ramp. Use key next to arrow, >, to move
the brackets, [ ] ,around the selection
- Press NEXT
S 03
xxxx V
>
SELECT VOLTAGE SLOPE
RAPID
[
]
>
NEXT
>
- The first line displays the step number, slope and
voltage
- Select VHOURS (default)
- Enter time (volt hours)
(7 cm: 10,000; 11 cm: 20,000; 17 cm 40,000)
- Press NEXT
S 03
xxxx V
>
HRS:MIN
VHOURS
ENTER TIME:
xxxxx
>
NEXT
>
10.