beautypg.com

Bio-Rad Mini-PROTEAN® 3 Dodeca™ Cell User Manual

Page 3

background image

Section 1
General Information

1.1 Introduction

The Mini-PROTEAN 3 Dodeca Cell is a multi-cell for mini-vertical gel electrophoresis.

It can run four to twelve identical polyacrylamide gels simultaneously. The Dodeca cell
includes clamping frames, buffer dams, a drain line, and gel releasers. Ready Gel precast gels
are compatible with the Mini-PROTEAN 3 Dodeca Cell. Handcasting electrophoresis gels is
also an option; see Section 5 for recommended handcast accessories.

Specifications

Tank and lid

Acrylic

Clamping frame

Polycarbonate and liquid crystal polymer

Upper electrode holder

Polycarbonate with 109 mm (43”) platinum wire

Lower electrode assembly

Polycarbonate with 89 cm (35”) platinum wire

Drain line

Tygon tubing

Drain line connectors

Delrin

Cooling coil

Acrylic

Cooling coil connector tubing

Tygon

Maximum buffer volume

4.4 liters

Minimum buffer volume

3.4 liters

Overall size

41.5 x 15 x 16.2 cm (L x W x H)

Precast gel compatibility

Ready Gel precast gels

Handcast gel compatibility

Mini-PROTEAN 3 glass plates and spacers

Safety limits

500V, 500W

Weight

5 kg (11 lb)

Note: Dodeca cell components are not compatible with acetone or ethanol. Use of organic
solvents voids all warranties. The Dodeca cell is compatible only with aqueous reagents.

1.2 Safety

Power to the Mini-PROTEAN 3 Dodeca cell is supplied by an external DC voltage power

supply (not included). The output of the power supply must be isolated from external ground
to insure that the DC voltage output floats with respect to ground. All Bio-Rad power supplies
meet this important safety requirement. Regardless of the power supply used, the maximum
specified operating parameters for the Mini-PROTEAN 3 Dodeca cell are as follows:

500 VDC

maximum voltage limit

500 W

maximum power limit

40 °C

maximum ambient temperature limit

The current to the cell enters the unit through the lid assembly and provides a safety

interlock to the user. The circuit between the cell and the lid’s upper electrodes is broken
when the lid is removed. Always turn off the power supply before removing the lid. Do not
attempt to use the cell without the safety lid.

1