beautypg.com

Maintenance 8” models, cont, Impellar and flow tube assemblies, Inlet spider – Baseline Systems BHM Series Hydrometer User Manual

Page 38

background image

38

Hydrometers Operation and Maintenance Manual

Maintenance 8” Models, Cont.

IMPELLAR AND FLOW TUBE ASSEMBLIES

Dis-assembly

• Remove valve seat base (49) from atop the flow tube (55). Be careful not

to damage the impeller shaft. Inspect the valve seat base for excessive

wear and replace as necessary.

• Remove the impeller (50).

• Inspect for cracks or excessive wear and check that the impeller shaft is

perfectly straight. Replace if necessary.

• Remove the flow tube (55).

• Inspect and replace as necessary.

• Inspect the upper and lower flow tube o-rings (54) and replace as

necessary.

• Remove the inlet spider (57) assembly. Inspect and repair as necessary.

Re-assembly

• Replace the inlet spider assembly (57) into the hydrometer body.

• Place the two lower flow tube o-rings (54) onto the flow tube (55).

• Place the upper flow tube o-ring (54) into the valve seat base (49).

• Replace the flow tube (55) atop the inlet spider assembly (57) in the

hydrometer body.

• Replace the impeller assembly (51) into the flow tube (55) - the impeller

shaft should rest in the lower bearing bushing (56).

• Place the valve seat base (49) over the impeller shaft so that it rests atop

the flow tube (55).

INLET SPIDER

Dis-assembly

• Unscrew and remove the cap (60). Inspect and replace as necessary.

• Remove the lock nut (59).

• Unscrew and remove the lower bearing screw (58) from the inlet spider

(57). Inspect and replace as necessary

• Inspect the inlet spider (57) and the lower bearing bushing (56) and

replace as necessary.

Re-assembly

• Screw the upper bearing bushing (56) back into the top of the inlet

spider (57).

• Replace the lower bearing screw (58) into bottom of the inlet spider.

Tighten approximately eight turns.

• Replace the lock nut (59) and the cap (60).