Device Placement (Bottom View)
NOTE:
1.
Two electrical/communication/medical gas devices cannot be placed together at any corner.
2.
Maximum of 12 gas outlets are allowed on the column. If 12 gas outlets are required, then side mount is optional.
3.
All electrical devices must be located within section A, B, C, or D only.
4.
The Nitrogen Control Panel will occupy three spaces of the bulkhead (E.G. A2 to A4 see Outline).
Location of Nitrogen Control Panel
Side A
Side B
Side C
Side D
Please Specify Outlet Language
US
CAN
FRENCH
SPANISH/US
SPANISH/ISO
Gas Legend
Oxygen
= O
Medical Air
= A
Vacuum
= V
Nitrous Oxide = 2
Nitrogen
= N
Carbon Dioxide = C
WAGD (US)
= W
AGSS (Canada) = E
Nitrogen Control Panel
Yes
o
No
Please fill the devices that you require for your application.
(Specify by using the symbols on Page 5 of this Approval Booklet.)
Approval Signature
Date
Phone No.
* if YES, please provide the location (ex. B1)
Is a Local Switch Required?
Yes
o
No
6
Amico Pipeline
- Artwall (6 pages)
- Ceiling Column Manual Retractable (16 pages)
- Ceiling Column Rigid (Stationary) (12 pages)
- Ceiling Column Rigid LITE (Stationary) (12 pages)
- Recessed Console Unit (8 pages)
- Freedom Series Recessed (12 pages)
- Freedom Series Surface Mounted (8 pages)
- LDR Headwall (8 pages)
- Majestic Series Double Tier Recessed Headwall (6 pages)
- Majestic Series Double Tier Surface Mounted Headwall (8 pages)
- Majestic Series Single Tier Surface Mounted Headwall (6 pages)
- Majestic Series Single Tier Surface Mounted Headwall (8 pages)
- Majestic Series Single Tier Recessed Headwall (8 pages)
- Noble Custom Cabinet (7 pages)
- Power Column (7 pages)
- Regal Series Recessed Vertical Flatwall (8 pages)
- Regal Series Surface Mounted Vertical Flatwall (7 pages)
- Security Recessed Console Box Assembly (12 pages)
- Surface Mount Console Unit (6 pages)
- Hydraulic Patient Transfer Stretcher (S-H-300) (20 pages)
- Hydraulic Patient Transfer Stretcher (S-H-300) (11 pages)
- AHM Monitor Arms (Adjustable Height) (34 pages)
- Barcode Dock Mount (15 pages)
- Condor Adjustable Height Channel (20 pages)
- Condor Lite Fixed Height Channel (31 pages)
- Condor Lite Fixed Height Channel (12 pages)
- Falcon Combo Arm (44 pages)
- Light-Duty Monitor Arm (20 pages)
- Monitor Roll Stands (12 pages)
- CSA Duplex RED Horizontal Tank Mount (36 pages)
- NFPA Duplex CCD Horizontal Tank Mount (51 pages)
- NFPA Duplex RED Horizontal Tank Mount (40 pages)
- 12" Floor Pedestal (8 pages)
- 2 in 1 Alarm System (56 pages)
- Alarm Valve Combo Unit (32 pages)
- Alert-3 LCD Alarm (28 pages)
- Automatic Cryogenic Liquid Solenoid Manifold (32 pages)
- Ball Valve (8 pages)
- British Standard Outlet (18 pages)
- British Standard Outlet Slim (18 pages)
- Chemetron Compatible Ceiling Column Outlet (24 pages)
- Chemetron Retro-fit Gas Control Panel (4 pages)
- Dome Loaded Manifold CSA (44 pages)
- Dome Loaded Manifold Nema-4 NFPA (44 pages)