beautypg.com

Test procedure – LaMotte Dissolved Oxygen Water Quality User Manual

Page 8

background image

TEST PROCEDURE

6

9.

6.

Continue

titrating until

the blue color

disappears and

the solution

becomes

colorless.

Carefully remove the

Titrator and cap. Do

not to disturb the

Titrator plunger.

Add 8 drops of

Starch Indicator

Solution

(4170WT). The

sample should

turn blue.

8.

11.

10.

Cap the titration

tube. Insert the tip of

the Titrator into the

opening of the

titration tube cap.

Insert the tip of the

Titrator into the

opening of the

titration tube cap.

Slowly depress the
plunger to dispense the

titrating solution until

the yellow-brown

color changes to a very
pale yellow. Gently

swirl the tube during

the tiration to mix the

contents.

7.

12.

Read the test result directly

from the scale where the large

ring on the Titrator meets the

Titrator barrel. Record as ppm

Dissolved Oxygen. Each

minor division on the Titrator

scale equals 0.2 ppm.

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1.0

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1.0

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1.0

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1.0

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1.0

4.0

5.0

Result:

4.0 ppm