beautypg.com

Epa compliance – LaMotte Dissolved Oxygen Water Quality User Manual

Page 11

background image

EPA COMPLIANCE

9

9.

Continue adding Sodium
Thiosulfate, 0.025N until

the blue color disappears

and the solution is

colorless.

10.

Read the test result directly from the scale
where the large ring on the Titrator meets the

Titrator barrel. Include the value of the original

amount dispensed (1 mL). If the reading is 2.0

+/-0.1 mL, the Sodium Thiosulfate, 0.025N
(4169) is satisfactory. If not, discard and

replace with new reagent.

While gently swirling the
tube, add Sodium

Thiosulfate, 0.025N until

the color fades to pale

yellow. It will be
necessary to refill the

Direct Reading Titrator.

7.

8.

Add 3 drops of
Starch

Indicator

Solution

(4170WT).
The solution

will turn blue.

2.0

3.0

Result:

2.0 ppm

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1.0

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1.0