Test procedure, Part 3 - the titration, Turn the bottle upright and remove the titrator – LaMotte Dissolved Oxygen Water Quality User Manual

Page 7: If the sample is a very pale yellow, go to step 9, Continued

background image

TEST PROCEDURE

PART 3 - THE TITRATION

5

Fill the titration

tube (0608) to the

20 mL line with the

fixed sample. Cap

the tube.

If small air bubbles appear in

the Titrator barrel, expel them

by partially filling the barrel

and pumping the titration

solution back into the reagent

container. Repeat until bubble

disappears.

NOTE:

2.

Depress plunger

of the Titrator

(0377).

Insert the Titrator

into the plug in the

top of the *Sodium

Thiosulfate, 0.025N

(4169) titrating

solution.

3.

5.

Turn the bottle

upright and remove

the Titrator.

NOTE:

If the sample is a very

pale yellow, go to Step 9.

4.

Invert the bottle and

slowly withdraw the

plunger until the

large ring on the

plunger is opposite

the zero (0) line on

the scale.

continued . . .