beautypg.com

LaMotte ATQ Digital Series Water Softener Demo Kit User Manual

Page 17

background image

17

5

+

3

Color

Q

6

Color

Q

5

* Reagent is a potential health hazard. Read MSDS at www.lamotte.com.

FREE / TOTAL CHLORINE

ColorQ Range: 0.0-10.0 ppm Chlorine

*Sulfi de Reagent A: Code V-4458LWT-G

*Sulfi de Reagent A: Code V-4458LWT-G

*Sulfi de Reagent A: Code V-4458LWT-G

1. Press button to go to “SuL” Sulfi de.

2. Fill a clean tube (0201) to the 5 mL line with water sample.

3. Add 5 drops of SULFIDE REAGENT A (4458).

4. Add 3 drops of SULFIDE REAGENT B (4459).

5. Cap and invert 3 times to mix. Allow tube to sit for ONE minute.

6. Add 5 drops of SULFIDE REAGENT C (4460).

7. Cap. Invert 3 times to mix. Insert tube into ColorQ.

8. Press button to read “SuL” Total Sulfi de in ppm. Remove tube.

Samples with sulfi de concentrations above 3 ppm must be diluted. The meter
will display “Hi”.

A. Use the 1 mL pipet (0354) to add 1 mL water sample to a test tube.

B. Dilute to the 5 mL line with distilled or deionized water. Cap and mix.

C. Blank with the diluted sample from Step B. Follow the test procedure,

steps 3 - 8, with the diluted sample. Multiply the displayed result by 5.

DILUTION