LaMotte TDS 5 Meter User Manual
Page 7
18
TDS 5 METER
TDS 5 METER
7
Meter Specifications
Range:
0.0 to 99.9 ppm
0.0 to 50°C
100 to 999 ppm
1.00 to 9.99 ppt
Auto Ranging
Resolution:
0.1 ppm
0.1°C
1 ppm
0.01 ppt
Accuracy:
+/- 2 ppm
+/- 0.5°C
+/- 20 ppm
+/- 0.2 ppt
(+/- 2% of full scale)
Calibration:
Up to 3 points
offset 0.1°C increments
TDS Factor:
User adjustable 0.5 to 0.85
Cell Constant:
1
Auto Shutoff:
after 17 minutes
Automatic Temp.
Comp.:
0.0 to 50°C
Reference Temperature: Factory set at 25°C
Temperature Coefficient: Factory set at 2% per °C
Keys:
ON/OFF, HOLD/ENTER, MODE/INC, CAL,
s/t
Power:
4 AAA Alkaline Batteries
Accessories
Description
Order #
TDS/Temp Probe
5-0042
Conductivity Standard, 52 ppm TDS, 500 mL
6416-L
Conductivity Standard, 503 ppm TDS, 500 mL
6417-L
Conductivity Standard, 989 ppm TDS, 500 mL
6354-L
Conductivity Standard, 4668 ppm TDS, 500 mL
6418-L
Protective Rubber Boot
5-0040
Error Messages
Error
Message
Indicates
Solution
E 2
Microprocessor
Error
Recalibrate instrument
E 3
Microprocessor
Error
Turn meter off and then on
again. Contact LaMotte for
return authorization
Ur
Temperature under range Warm sample
Faulty sensor
Contact LaMotte for return
authorization
Or
TDS over range
Dilute sample
Temperature over range
Cool sample
Faulty temperature sensor Contact LaMotte for return
authorization