LaMotte TDS 5 Meter User Manual
Page 10
10
TDS 5 METER
TDS 5 METER
15
TDS Calibration
The meter is capable of up to a three-point calibration. Only one
calibration is needed for the entire range of the meter but better accuracy
will be achieved if each range is calibrated. If a range is not calibrated, the
meter will automatically calibrate to the range closest to the sample being
tested.
Calibrate the meter at least once a week if solutions with a conductivity of
less than 50 ppm are being measured in order to get +/- 2% full scale
accuracy. If solutions are being measured with conductivity in the mid
range, and the probe is washed in deionized water and stored dry, the meter
need only be calibrated once a month. Calibrate the meter at least once a
week if measurements are taken at extreme temperatures.
Select a TDS standard with a concentration near the sample to be
measured. If this is not possible, choose a value that is approximately 2/3 of
the full scale value of the measurement range. For example, in the 100 to
999 ppm range, the meter should be calibrated with a 989 ppm solution.
NOTE: The meter will not accept calibration values less than 20 ppm. All new
calibrations will automatically override existing calibrations.
3.
1.
2.
4.
Press the ON/OFF
key to turn the
meter on and to
enter the TDS
mode.
1
2
Pour approximately 3 cm of a
conductivity standard into two
separate containers.
Conductivity
Standard
3 cm
1
2
Gently stir the first
container of the
standard with the
probe. This
container will be
used as the rinse
solution to remove
contaminants.
1
2
Immerse the
probe in the
second
container of the
standard.
CAL
MODE
HOLD
ENTER
ON
OFF
Measuring the TDS of Soil
continued...