Ho lo °c pptm – LaMotte TDS 5 Meter User Manual
Page 6
6
TDS 5 METER
TDS 5 METER
19
ON/OFF :
Turns the meter on and off
HOLD/ENTER:
Freezes the measured reading or confirms the calibration
value.
CAL:
Allows calibration of the meter for TDS and temperature
MODE/INC:
Selects the parameter to be measured: TDS or temperature
: Scrolls display values up or
down to select
calibration values
HO
LO
°C
pptm
1.8.8.8
indicates
HOLD
function
is activated
indicates
low
battery
condition
indicates
operation mode
Degree C annunciator
ppm, ppt annunciators
for TDS readings
Reading
µS, mS
annunciator
for conductivity
readings
mµS
Display and Keypad Functions
ON
OFF
CAL
MODE
HOLD
ENTER
Problem
Check
Action
LO Message
Batteries
Replace
Power on but no display
Batteries
Insert batteries
Batteries
Verify correct polarity
Batteries
Replace
Unstable reading
Probe
Immerse probe more deeply in
sample
Probe
Replace probe
Interference
from nearby
motor
Remove or switch off
Probe
Clean probe
Slow response
Probe
Clean probe
Display locked
Electrical
discharge
near meter
Switch meter off and then on
again. Remove batteries and
reinsert.
Unable to calibrate
Probe
Clean probe
Calibration
standard
Select standard within range
Troubleshooting