beautypg.com

LaMotte Tracer SAL Meter User Manual

Page 9

background image

Measurement
1. Fill a sample cup to the 20 mL line with the test sample. Sample

depth must be greater than or equal to 1.5 inches.

2. Immerse the TRACER electrode in the sample. Make sure the

electrode is completely submersed.

3. Press the ON/OFF button. (8888 and then

SELF CAL will appear in the display during the
initial diagnostics).

4. Press and hold the MODE/HOLD button to scroll

to the desired measurement mode.

5. Slowly stir the sample with the TRACER to

remove air bubbles.

6. The meter will autorange to the proper range

and the reading will be displayed.

7. Rinse the electrode in distilled water. Replace

the cap.

Measuring the TDS of Soil
1. Fill a 50 mL beaker with the soil sample. Tap the beaker lightly on a

hard surface to remove trapped air. Remove excess soil from the
surface.

2. Empty the soil into a 250 mL wide-mouth flask.
3. Add 100 mL of distilled water. Stopper and shake vigorously.
4. Wait 30 minute. (Shake the flask vigorously three or four times

during this period.)

5. Filter the contents of the flask. Collect the filtrate in a beaker.
6. Rinse the electrode with distilled or deionized water to remove

impurities.

7. Press the ON/OFF button to turn the TRACER on. Make sure the

meter is in the TDS mode.

8. Immerse the electrode in the filtrate. Make sure the tip of the

electrode is completely immersed.

9. Stir the filtrate with the electrode to create a homogeneous solution.

9

This manual is related to the following products: