LaMotte Tracer SAL Meter User Manual
Page 16
Electrode Cleaning Recommendations
Do not soak the electrode in the solutions for longer than the
recommended length of time. To do so may cause a reference
potential shift which will cause a degradation in performance or
failure. When cleaning the probe, do not scratch or damage the
platinized electrode surfaces.
Contaminant
Cleaning Solution
Procedure
Water soluble
substances
Deionized water
Soak or scrub with a soft brush.
Rinse thoroughly with DI
water and dry.
Grease and
oil
Warm water and
household detergant
Soak or scrub with a soft brush,
maximun 10 minutes. Rinse
thoroughly with DI water and
dry.
Heavy grease
Alcohol
Soak for a maximum of 5
minutes. Scrub with a soft
brush. Rinse thoroughly with
DI water and dry.
Lime and
hydroxide
coatings and oil
10% Acetic acid
Soak until coating dissolves,
maximum 5 minutes. Rinse
thoroughly with DI water and
dry.
16