beautypg.com

LaMotte Tracer SAL Meter User Manual

Page 10

background image

10. Gently stir the filtrate with the electrode. Read the TDS value of

the filtrate from the display.

11. Rinse the electrode in distilled water. Replace the cap.
Storing Readings
1. After the reading is displayed press and hold the MODE/HOLD

button to store the current reading. The meter will enter the HOLD
mode and HOLD will be displayed. The storage location number
will be displayed on the lower display followed by the reading being
stored.

2. Press the MODE/HOLD button to exit the HOLD mode and return

to normal operation.

3. If an attempt is made to store more than 25 readings, the stored

readings will be overwritten starting with the first reading.

Recalling Stored Readings
NOTE: First ensure that the HOLD symbol is not displayed. If it is, exit
the HOLD function by pressing the MODE/HOLD button.
1. Press the CAL/RECALL button and then press the MODE/HOLD

button; the location number (1 through 25) will briefly appear and
then the value stored in that location will appear. The displayed
units will flash, indicating that the storage recall mode is active.

2. The last stored reading taken

will be displayed first. To
advance to the previously
stored readings, press the
MODE/HOLD button. The
location number is displayed
first, followed by the reading
stored in that location.

3. To exit the storage mode,

press the CAL/RECALL
button and the
TRACER will return to
normal operation after
displaying “End”.

10

EC/TDS/SAL

ON

OFF

MODE

CAL

100.9

72.3

°F

Conductivity

TDS (ppm)

m

PP

Salinity

m

PP

4.01

This manual is related to the following products: