LaMotte 2020i Turbidimeter User Manual
Page 9
2. Scroll down and then press
*IOK
to select Options.
M a i n M e n u
M e a s u r e
D a t a L o g g i n g
* O p t i o n s
1 6 : 0 2 : 1 9
0 1 / 0 4 / 0 5
3. Scroll down and then press
*IOK
to select Turbidity.
O p t i o n s
A v e r a g i n g
* T u r b i d i t y
D a t e / T i m e
L a n g u a g e
1 6 : 0 2 : 1 9
0 1 / 0 4 / 0 5
4. Press the
*IOK to select Units.
T u r b i d i t y
* U n i t s
C a l i b r a t i o n
1 6 : 0 2 : 1 9
0 1 / 0 4 / 0 5
5. Scroll down and then press
*IOK
to select units.
Available units are: NTU
(Nephelometric Turbidity Units);
FNU
(Formazin Nephelometric
Units); ASBC (American Society of
Brewing Chemists); EBC (European
Brewery Convention)
U n i t s
N T U
F N U
A S B C
E B C
1 6 : 0 2 : 1 9
0 1 / 0 4 / 0 5
Note: If Attenuation is chosen as a calibration curve. The result will be reported in
FAU (Formazin Attenuation Units).
6. Press
to exit to a previous menu or
make another menu selection or
press
OFF to turn the meter off.
OPTIONS & SETUP • TURBIDITY UNITS
9