LaMotte 2020i Turbidimeter User Manual
Page 10
SELECTING A TURBIDITY CALIBRATION CURVE
1. Press
ON to turn the meter on.
1 . 3
2. Scroll down and then press
*IOK to
select Options.
M a i n M e n u
M e a s u r e
D a t a L o g g i n g
* O p t i o n s
1 6 : 0 2 : 1 9
0 1 / 0 4 / 0 5
3. Scroll down and then press
*IOK
to select Turbidity.
O p t i o n s
A v e r a g i n g
* T u r b i d i t y
D a t e / T i m e
L a n g u a g e
1 6 : 0 2 : 1 9
0 1 / 0 4 / 0 5
4. Scroll down and then press
*IOK
to select Calibration.
T u r b i d i t y
U n i t s
* C a l i b r a t i o n
1 6 : 0 2 : 1 9
0 1 / 0 4 / 0 5
5. Scroll down and then press
*IOK
to select a Calibration curve. Select
a calibration option based on the
composition of the standards that
will be used to calibrate the meter.
Available options are: Formazin,
Japan Standard
, Attenuation.
T u r b i d i t y
U n i t s
* C a l i b r a t i o n
1 6 : 0 2 : 1 9
0 1 / 0 4 / 0 5
For the most accurate results, the
Attenuation option should be chosen
when samples are over 500 NTU.
The range for the Attenuation option
is 40–4000 NTU.
C a l i b r a t i o n
* F o r m a z i n
J a p a n S t a n d a r d
A t t e n u a t i o n
1 6 : 0 2 : 1 9
0 1 / 0 4 / 0 5
10
OPTIONS & SETUP • TURBIDITY CALIBRATION