beautypg.com

LaMotte 2020i Turbidimeter User Manual

Page 10

background image

SELECTING A TURBIDITY CALIBRATION CURVE

1. Press

ON to turn the meter on.

1 . 3

2. Scroll down and then press

*IOK to

select Options.

M a i n M e n u

M e a s u r e
D a t a L o g g i n g

* O p t i o n s

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

3. Scroll down and then press

*IOK

to select Turbidity.

O p t i o n s

A v e r a g i n g

* T u r b i d i t y

D a t e / T i m e
L a n g u a g e

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

4. Scroll down and then press

*IOK

to select Calibration.

T u r b i d i t y

U n i t s

* C a l i b r a t i o n

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

5. Scroll down and then press

*IOK

to select a Calibration curve. Select
a calibration option based on the
composition of the standards that
will be used to calibrate the meter.
Available options are: Formazin,
Japan Standard

, Attenuation.

T u r b i d i t y

U n i t s

* C a l i b r a t i o n

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

For the most accurate results, the
Attenuation option should be chosen
when samples are over 500 NTU.
The range for the Attenuation option
is 40–4000 NTU.

C a l i b r a t i o n
* F o r m a z i n

J a p a n S t a n d a r d
A t t e n u a t i o n

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

10

OPTIONS & SETUP • TURBIDITY CALIBRATION

This manual is related to the following products: