°c

°f

coRRection factoR

30

86

1.16

29

84.2

1.13

28

82.4

1.09

27

80.6

1.06

26

78.8

1.03

25

77

1.00

24

75.2

0.97

23

73.4

0.94

22

71.6

0.92

21

69.8

0.89

20

68

0.86

19

66.2

0.84

18

64.4

0.81

17

62.6

0.79

16

60.8

0.77

15

59

0.74

14

57.2

0.72

13

55.4

0.70

12

53.6

0.68

11

51.8

0.66

10

50

0.64

9

48.2

0.62

8

46.4

0.61

7

44.6

0.59

6

42.8

0.57

5

41

0.55

Multiply the nominal product flow at 25° C by the temperature correction factor to

determine the flow at various other temperatures.

VI. Appendix

The following tables are intended as a guide to determining the flow

rates for the PWR4024 series RO systems. All flows are in gallons

per minute (GPM).
Nominal flows for systems with reject recycle and a feed water

Silt Density Index less than 3.

Nominal flows for systems with reject recycle and a feed water

Silt Density Index of 3 to less than 5.

model numbeR

PWR40243163

PWR40243203

PWR40243243

Product (max)

20

25

30

Reject

6.7

8.4

10

Reject Recycle

4.1

5.1

6.2

model numbeR

PWR40243163

PWR40243203

PWR40243243

Product (max)

18

22.5

27

Reject

6

7.5

9

Reject Recycle

3.7

4.6

5.5

Temperature Correction Factors