Metrohm Inline Dialysis User Manual
Page 13

■■■■■■■■■■■■■■■■■■■■■■
3 Installation
IC equipment: Dialysis
■■■■■■■■
7
■
Loosen the screws with the hex key.
■
Remove the screws with the washers and put them aside.
3 Disassembling the dialysis cell
■
Remove the upper part of the dialysis cell.
■
Remove the sealing ring.
4 Cleaning the dialysis cell
CAUTION
Damage to the dialysis cell
Organic solvents (e.g.: acetone) corrode and damage the dialysis
cell material (PMMA).
■
Use only ultrapure water or a water-ethanol mixture (70:30) for
cleaning the dialysis cell.
■
For samples that contain organic components (e.g. solvents),
use the PEEK dialysis cell (6.2729.120). This cell has an excellent
chemical resistance to organic chemicals.
■
Thoroughly rinse off the sealing ring (2-3), the left chamber (2-2)
and the right chamber (2-4) of the dialysis cell with ultrapure
water.
■
Dry all parts with a lint-free cloth.