Application Bulletin 353

Installation Instructions of TitrIC pro I

Page 14 of 14

TitrIC pro I_pH calibration manual:

Here the iAquatrode

is taken out of the titration vessel and placed into the
buffer solutions. Start this method only from tiamo

TM

!

TitrIC pro I_titer determination of HCl:

This method is

used for the titer determination. It has to be used in
combination with the MagIC Net method

TitrIC pro

I_MiPT titer determination of HCl

.

TitrIC pro I_titer determination of EDTA:

This method is

used for the titer determination. It has to be used in
combination with the MagIC Net method

TitrIC pro

I_MiPT titer determination of EDTA

.

TitrIC pro I_conductivity calibration:

This method is used

for calibration of the conductivity sensor. It has to be
used in combination with the MagIC Net method

TitrIC

pro I_MiPT calibration conductivity

.

Import: Click on

Method

at the left (in MagIC Net and

tiamo

TM

), then select under

File / Method manager… / Edit /

Import

Import the methods related to TitrIC pro I with the

MiPT kit from the MagIC Net CD. You will find the methods
in the folder example/TitrIC of the CD.

Note!

Alterations of these methods should only be carried out by a
person at the administrator level and who is thoroughly
familiar with tiamo

TM

and MagIC Net

With the TitrIC pro – Combo Kit Partial Loop following
analysis can be carried out using the TitrIC pro I_MiPT
methods in MagIC respectively tiamo

TM

:

The whole procedure is controlled by the MagIC Net
software which is the master software. In tiamo

TM

the

method

TitrIC pro I

is selected as a single determination.

Start the determination in autorepeat mode. To set it into
autorepeat go to the Run window, right-mouse-click and
select

Properties Run window

. Under Process select

Autorepeat determination

. Close the window with

. Start

the determination. Now tiamo

TM

is waiting for the start

signal from MagIC Net via RS232.

In the MagIC Net software select the method

TitrIC pro

I_MiPT

. Enter the required sample position, sample

identification etc. into MagIC Net. For value 1 choose
according to the list below:

Value 1 = 1 complete analysis including titration and
ion chromatography.

Value 1 = 2 only an ion chromatography analysis (e.g.
a calibration)

Value 1 = 3 only a titration analysis