6 installing the dosino, Installing the dosino – Metrohm 941 Eluent Production Module User Manual
Page 34

4.6 Installing the Dosino
■■■■■■■■■■■■■■■■■■■■■■
26
■■■■■■■■
941 Eluent Production Module (2.941.0010)
Installing an ELGA PURELAB flex
Install the ELGA PURELAB flex as described in the manufacturer's docu-
mentation.
Carry out the following steps before installing the composite aeration fil-
ter.
1 Installing the ultrapure water insert
Screw the ultrapure water insert supplied in the opening in which the
composite aeration filter is to be installed according to the manufac-
turer's documentation.
2 Inserting the composite aeration filter
Screw the composite aeration filter onto the ultrapure water insert.
3 Connecting the FEP tubing
Screw the FEP tubing onto one of the connectors of the ultrapure
water insert.
4.6
Installing the Dosino
Attaching the Dosino to the Dosing Unit
Accessories
For this step you need the following accessories:
■
800 Dosino (2.800.0010)
■
Dosing Unit 50 mL (6.3032.250)
The manual for the 800 Dosino describes how to attach the 800 Dosino
correctly to a dosing unit.
CAUTION
Please read through the correct procedure in the manual for the 800
Dosino before you attach the Dosino to the dosing unit.
Installing FEP tubings
Accessories
For this step you need the following accessories:
■
800 Dosino (2.800.0010) with dosing unit 50 mL (6.3032.250)
attached
■
The three FEP tubings connected to the bottles,
one additional FEP tubing (6.1805.530)