4.3 Connecting the eluent bottle
■■■■■■■■■■■■■■■■■■■■■■
16
■■■■■■■■
941 Eluent Production Module (2.941.0010)
A longer section of silicone tubing (6.1816.020) may have to be cut
to size and mounted (see also the manual for the IC instrument).
5 If one of the instruments in the stack is equipped with a leak sensor
connection socket, connect the leak sensor (see manual of the IC
instrument).
6 Restore any capillary connections that may have been removed.
4.3
Connecting the eluent bottle
For the automated eluent production, the eluent bottle must be equipped
with a bottle cap for a level sensor and a sensor of the type "FULL".
Inserting the measuring sensor
Accessories
For this step you need:
■
A bottle cap for level sensor (6.1626.000)
■
FULL sensor (6.2769.000) for anion eluents or EMPTY sensor
(6.2769.1X0) for cation eluents
Figure 3
Bottle cap for level sensor - Inserting the measuring sensor
1
Measuring sensor opening
For two measuring sensors (inside) and the
level sensor cable (outside).
2
M6 opening
For the FEP tubing connection to the Dos-
ino.
3
Adsorber opening
4
M8 opening
For the capillary connection to the ion chro-
matograph.
1 Remove the stoppers from the bottle cap.
2 Place the bottle cap on a surface with the opening facing up.