5 miscellaneous – Metrohm TitrIC 2 User Manual
Page 11
Application Bulletin 286 e
Installation Instructions of TitrIC 2
Page 11/16
5 Miscellaneous
5.1 Stirrer
The 802 Rod Stirrer is inserted through opening 8 of the external cell and connected to the tower of the
855. The propeller (6.1909.020) is pushed on from below.
5.2 Length of the aspiration tips in the external cell
Clamp the titration vessel to the titration vessel lid.
The M8 aspiration tip in opening 1 must reach right to the base of the titration vessel as this is
the only way to ensure that the external cell is cleaned properly.
The buret tips in openings 3 and 4 should reach the middle of the titration vessel.
5.3 Electrodes
The pH electrode (Aquatrode Plus with Pt 1000; 6.0257.000) is inserted into opening 9 together
with the SGJ sleeve (6.1236.050). The cable is connected to the 855 Robotic Titrosampler at
the rear top right. The plug of the gray cable is connected to the opening marked
Ind.
. The red
plugs with the adapters 6.2103.140 and 6.2103.130 are inserted in the openings marked
Temp.
If the electrode is not to be used for some time then it should be returned to its holder and kept
in the storage solution (6.2323.000).
Conductivity cell (6.0912.110): The measuring cell cable is led through the guide chain of the
855 and the two black plugs are inserted into the openings marked
Cond. Cell
of 712
Conductometer, the two red plugs into those marked
Pt 100/1000
. Later (as soon as the titration
head has been mounted) the measuring cell with the SGJ sleeve (6.1236.050) will be inserted
in the SGJ 14 opening of the titration head.